Back to DriveX™

DriveX™ Manual

Warranty

INTENDED USE

This cordless electric dental torque driver is a motorized handpiece system designed for securing abutments onto dental implants during implant surgery. It must be used only under the supervision of qualified dental professionals who have been properly trained in its operation.

CONTRAINDICATIONS

This device is contraindicated for use in patients with active medical implants such as pacemakers, cochlear implants, or similar devices.
The safety and efficacy of this product have not been established for use in pregnant women or pediatric patients.

WARRANTY AND LIMITATION OF LIABILITY

IMPORTANT — READ CAREFULLY BEFORE USE

This product is intended strictly for professional dental use by trained and licensed healthcare providers. Conmetior warrants that the device is free from manufacturing defects in material and workmanship under normal use for a period of one (1) year from the date of purchase. This limited warranty applies only to the original purchaser and is not transferable.

Warranty Exclusions and Limitations

Conmetior shall not be held liable under any circumstances for the following, including but not limited to:

1. PATIENT INJURY OR DEATH

Under no condition shall Conmetior be liable for any injury, medical complication, or death of a patient resulting from the use, misuse, or malfunction of this device. Clinical risks and outcomes are the sole responsibility of the licensed practitioner using the device.

2. OPERATOR INJURY OR NEGLIGENCE

Any use of this device in a manner not expressly described in the instructions, including lack of attention, user error, or failure to follow accepted clinical protocols, voids all warranties and releases Conmetior from any and all liability.

3. ACCIDENTAL DAMAGE OR DROPPING THE DEVICE

This device is a precision instrument. Dropping the unit or any of its components — whether or not it appears to function afterward — voids the warranty. Damage caused by impact, mishandling, or improper storage is not covered and is the sole responsibility of the purchaser.

4. IMPROPER CLEANING & STERILIZATION

Damage caused by failure to follow sterilization guidelines, use of unapproved disinfectants, ultrasonic cleaners, or exposure to excessive heat or moisture will void the warranty.

5. USE BY UNQUALIFIED PERSONNEL

Use of the device by individuals who are not trained, licensed, or otherwise qualified to perform dental procedures absolves Conmetior of all liability.

6. UNAPPROVED OR OFF-LABEL USE

The device must only be used for the purposes explicitly stated in this manual. Any use beyond these indications is considered off-label and is undertaken entirely at the user’s own risk.

7. ELECTROMAGNETIC INTERFERENCE OR ENVIRONMENTAL FACTORS

Malfunction due to failure to follow electromagnetic compatibility (EMC) guidelines or recommended environmental/storage conditions is not covered under warranty.

8. INDIRECT, INCIDENTAL, OR CONSQUENTIAL DAMAGES

Conmetior shall not be liable for any indirect, incidental, special, or consequential damages, including but not limited to lost income, clinical delays, business interruption, or legal claims stemming from use of this device.

Legal Disclaimer of Responsibility

By purchasing, opening, or using this product, the purchaser/user explicitly agrees to all terms stated in this manual and acknowledges full responsibility for the safe handling and clinical use of the device.

The purchaser understands and accepts that:

  • This device is not guaranteed to prevent surgical or clinical complications.

  • All responsibility for patient safety rests with the licensed operator.

  • Conmetior is not liable for injury, death, or damage resulting from use of this product under any circumstance.

If you do not agree to these terms, return the unused product in its original packaging for a full refund.

Part Identification

  1. Handpiece (1 unit)

  2. Spray nozzle (1 unit)

  3. Contra-angle attachment (1 unit)

  4. Adapter (1 unit)

  5. Charging base (1 unit)

Technical Specifications

  • Electrical Classification:

    • Protection type: Class II

    • Protection degree: Type B

  • Battery:

    • Type: Rechargeable lithium-ion

    • Model: 18500

    • Capacity: 1900 mAh

    • Safety Features: Over-voltage, over-current, and short-circuit protection

  • Adapter:

    • Input: 100–240V AC, 50/60 Hz

    • Output: 5V DC, 1A

  • Torque Range: 10–50 Ncm

  • Speed Range: 10–50 RPM

  • Contra-Angle Ratio: 20:1

  • Operation Modes: Forward (clockwise), Reverse (counter-clockwise)

Operating Conditions

  • Intended Use Location: Enclosed clinical environments

  • Ambient Temperature: 5°C to 40°C

  • Relative Humidity: <80%

  • Altitude: <2000 meters

  • Atmospheric Pressure: 70–106 kPa

Transport and Storage Conditions

  • Temperature: -20°C to +55°C

  • Relative Humidity: 20% to 80%

  • Atmospheric Pressure: 70–106 kPa

User Interface 

  1. Main Switch to power the device

  2. Display screen

  3. Direction Button to change direction

  4. S button to change speed

  5. N button to change torque

Charging The Unit

Connect the power adapter to the charging port located at the bottom of the handpiece. Ensure the connection is secure before charging.

While the device is charging, the battery icon on the screen will display a sequence of symbols to indicate charging progress, as shown below:

Low Battery Indicator

When the battery power is low, the following icon will appear on the screen to alert the user:      

A full charge typically takes up to five hours, depending on the remaining battery level and the battery's condition (state of health).

Once charging is complete, the battery icon on the screen will display as shown below:

⚠️ Warning

  • Before first use, charge the device for at least 5 hours to ensure optimal battery performance.

  • Only use the original Conmetior adapter and battery supplied with the device. Use of unauthorized components may result in malfunction or damage.    

Installing the contra-angle

Plug the contra-angle into the handpiece until it is securely seated. You will hear a loud “click”, indicating it is locked into place.

The contra-angle features a tri-lobed male connector, which fits precisely into the six-slot interface of the handpiece. Once engaged, the contra-angle head can rotate 360 degrees, allowing for optimal positioning during use.

Operating the Device

Powering on the Device

Press “S” button to change the Speed setting

Press “N” button to change the Torque setting

1. Working Mode Selection

Auto Mode (Default on First Use)

The device starts in Auto Mode by default. In this mode:

  • The startup data is the same as the last session.

⚠️ Make sure to verify or adjust settings when using a different implant system to ensure compatibility with the manufacturer’s specifications.

  • The device will continue operating until:

    • The power button is pressed again, or

    • The preset torque limit is reached.

  • Manual Mode

  1. Press and hold the power button to start the unit.

  2. Simultaneously press the “N” button and the button.

  3. A hand icon will appear on either the green (forward) or purple (reverse) arrow, indicating Manual Mode is active.

In this mode:

  • The device will only operate while the power button is being held down.

  • Releasing the power button will stop the motor immediately.

2. Switching Modes

To toggle between Auto Mode and Manual Mode:

  1. Press the “N” and button to switch back to auto mode.

  2. The mode icon will switch between Auto and Manual, confirming the current setting.

  3. Repeat the same button combination to switch back if needed.

3. Powering Off

To turn off the device:

  1. Press and hold the “N” key.

  2. While still holding “N,” press the power button once.

  3. After a “beep-beep” confirmation sound, the LCD screen will shut off.

  4. All functions will be disabled, and the main unit will be fully powered down.

4. Torque Holding Function (Manual Mode Only)

This feature is available only in Manual Mode with forward rotation:

  • When the preset torque is reached, continue holding the power button to maintain the torque output.

  • The device will maintain the torque for up to 15 seconds, after which it will automatically stop.

  • The user may also release the button manually to stop at any time.

5. Forward/Reverse Operation

  • Toggle between Forward Mode (▶ green arrow) and Reverse Mode (◀ purple arrow) based on clinical needs.

Forward Mode

  • Speed and torque are fully adjustable to suit the procedure.

Reverse Mode

  • Torque is automatically set to 50 Ncm.

  • The display will show the reverse torque icon.

  • The device will emit a continuous “beep-beep-beep” sound as a safety alert.

⚠️ Important: In Reverse Mode, the motor will not stop automatically. You must press or release the power button manually to stop the device.

Customizing user settings

Refer to booklet

DRAWINGS

Maintenance & Sterilization

To ensure proper hygiene and patient safety, all components must be thoroughly cleaned, disinfected, and sterilized before the first use and after each subsequent use.

Important Guidelines

  • Follow national and local sterilization regulations.

  • Use only disinfectants approved for medical devices (VAH/DGHM listed, CE marked, FDA-approved).

  • Do not lubricate the motor handpiece.

  • Do not clean the contra-angle in an ultrasonic cleaner.

  • Do not use bleach or chlorine-based disinfectants.

  • Use clean, demineralized water for the final rinse.

The user is solely responsible for ensuring sterility prior to each use. Always wear appropriate personal protective equipment (PPE) during handling and maintenance.

Autoclavable components

DRAWING

DRAWING

DRAWING

Warning

Only the components listed above are suitable for autoclaving.
Before initial use and after each subsequent use, these components must be sterilized using steam sterilization..

CLEANING AND DISINFECTING

1 Preparation

Disconnect all components from the motor handpiece. Store instruments in a dry, clean environment — not in humid surroundings.

2 Transportation

Ensure safe and protective storage of all components during transport to avoid damage or contamination.

3 Pre-Decontamination

Disassemble the device completely prior to cleaning. Only disassembled parts should be reprocessed.

4 Manual Pre-Cleaning

Visually inspect and manually clean all components. Submerge them in cleaning solution and flush lumens with a cold water jet for at least 10 seconds. Use a soft-bristle brush for external surfaces.

5 Automated Cleaning

Preferred method due to consistency and safety. Place parts in a washer-disinfector tray and run the following cycle: • 4 min pre-wash with cold water (<40°C) • 5 min wash with mild alkaline detergent at 55°C • 3 min neutralization with warm water (>40°C) • 5 min intermediate rinse with warm water (>40°C)

 6 Desinfection 

Use an automated washer-disinfector with a validated thermal disinfection cycle. The recommended cycle includes:

5 minutes at 93°C (199.4°F)

This cycle has been validated to achieve high-level microbial kill, equivalent to industry-recognized disinfection standards. Use only FDA-cleared washer-disinfectors and ensure all cycle parameters meet CDC guidelines for reprocessing reusable dental instruments.

Note: All components must be placed in the washer-disinfector in a fully disassembled state. Follow manufacturer instructions for positioning to ensure complete exposure to heat and detergent.

7 Drying

Use automatic dryer cycle in the washer/disinfector. If needed, manually dry with a lint-free towel. Use sterile compressed air to dry internal cavities.

8 Maintenance

Inspect all components for damage or wear. Clean and disinfect any visibly soiled parts again. Lubricate the contra-angle as instructed.

9 Packaging

Use approved sterilization packaging to prepare clean and dry parts for autoclaving.

10 Sterilization

Perform steam sterilization at 135°C for at least 4 minutes. Allow 10 minutes minimum drying time post-cycle.

11 Storage

Store components in their sterilization packaging in a dry, clean, and dust-free environment until next use.

Use only FDA-cleared steam sterilizers (autoclaves) that are properly maintained and validated for use with dental instruments.
Before initial use and after every procedure, all sterilizable components must undergo steam sterilization.

Allow all components to cool completely before handling.
Before use, inspect the sterilization packaging to ensure it is intact, dry, and within its expiration date.
If the packaging is damaged, wet, or expired, the components must be repackaged and sterilized again.

Note: Disinfection alone is not sufficient. Steam sterilization is required before the first use and after every use.

Disinfection of other components

DRAWING

DRAWING

DRAWING

Disinfection of External Surfaces

Wipe all external surfaces using a lint-free cloth lightly moistened with ethanol (70–80% by volume).
Surfaces must remain visibly wet for at least 2 minutes.
Repeat the wiping process five times to ensure effective disinfection.

Important: Only ethanol should be used for surface disinfection. Do not use other disinfectants, sprays, or chemical wipes, as they will damage the device.

Battery Maintenance

  • Charge the battery when the battery icon flashes on the display.

  • If the device is not used for an extended period, charge it fully at least once per month to prevent deep discharge and maintain battery health.

Troubleshooting

Refer to the troubleshooting section in the full manual for common issues and their solutions. If problems persist, do not attempt repair. Contact Conmetior for authorized support.

IssuePossible CauseSolution
No display / No response• Battery is fully discharged • Battery protection mode is active • Faulty battery• Charge the battery • Activate battery by charging • Contact Conmetior for assistance
Device cannot be charged• Adapter not properly connected • Incorrect adapter used • Faulty adapter• Reconnect the adapter securely • Use the original Conmetior adapter • Contact Conmetior
Battery runtime is noticeably reduced• Battery capacity has degraded• Replace with a new battery • Contact Conmetior for support
Alarm sound when tool is powered on• Tool is in counterclockwise (CCW) motion mode• Stop the motor • Switch to clockwise (CW) motion
No display / No responseBattery is out of power / Battrey is protected Faulty battery• Charging • Charging activation battery • Please contact Conmetior
The tool cannot be charged• The adapter is not well connected • Using the wrong adapter • Faulty adapter• Reconnect the adapter • Use the original adapter • Please contact Conmetior
Effective duration of the battery becomes shortThe capacity of the battery decreased.• Replace the battery with a new one • Please contact Conmetior.
There are an alarm sounds after starting the toolThe alarm beep sound indicates that the tool is under CCW motion.Stop the motor of the tool and change it to CW motion

Symbols

Environmental Disposal

  • Packaging materials: Recyclable

  • Metal components: Dispose as scrap metal

  • Electrical parts and PCBs: Dispose as electronic waste

  • Lithium batteries: Treat as hazardous waste per local laws

Manufacturer Rights

Conmetior reserves the right to modify product specifications, design, and documentation without prior notice.
Images included in this manual are for reference only.
All product designs, including internal architecture and aesthetic form, are protected by patent and intellectual property laws. Unauthorized reproduction or imitation is strictly prohibited.